Type-N Connector Assembly; Emoval - HP 8753D Option 011 Service Manual

Network analyzer
Hide thumbs Also See for 8753D Option 011:
Table of Contents

Advertisement

Type-N Connector Assembly

Tools Required
T-10 TO X screwdriver
T-15 TO X screwdriver
small slot screwdriver
ESD (electrostatic discharge) grounding wrist strap
5/16-inch open-end torque wrench (set to 10 in-lb)
Removal
1. Disconnect the power cord and remove the top cover (refer to \Covers" in this chapter).
2. emove the front panel (refer to \Front Panel Assembly" in this chapter).
3. emove the source bracket (item 1).
4. Disconnect the semi-rigid cables (W1, W2, W3, and W4) from the source and the samplers.
5. emove the two screws (item 2) that secure the bracket (item 3) holding the semi-rigid
cables against the card cage assembly. emove the bracket.
6. emove the right-side trim strip (item 4) from the front frame. emove the screw (item 5)
that secures the right end of the Type-N connector bracket.
14-16 Assembly Replacement and
Post-Repair Procedures