Advertisement

5. Write the following information on the disk label:
analyzer serial number
today's date
\EE ROM Backup Disk"
How to Retrieve Correction Constant Data from the EEPROM Backup Disk
1. Insert the \EE ROM Backup Disk" into the analyzer disk drive.
2. Make sure the A9 CC jumper is in the ALTER position.
3. ress
SAVE/RECALL 5
the le \N12345" where N12345 represents the le name of the EE ROM data for the
analyzer. On the factory shipped EE ROM backup disk, the lename is FILE1.
4. ress
RETURN RECALL TATE
instrument EE ROM's.
5. erform \Serial Number Correction Constants (Test 55)", and if applicable, \Option
Numbers Correction Constant (Test 56)."
6. ress
and verify that good data was transferred to EE ROM by performing a
PRESET 5
simple measurement.
7. Move the A9 CC jumper back to its NORMAL position when you are done working with
the instrument.
RAFT
3/21/106 15:11
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
ELECT DI K INTERNAL DI K
to down load the correction constants data into the
. Use the front panel knob to highlight
Adjustments and Correction Constants 3-39