Vertical osition and Focus Adjustments

Required Equipment and Tools
Torx screwdriver T-15
Non-conductive at head screwdriver,-2 inches
Antistatic wrist strap
Antistatic wrist strap cord
Static-control table mat and earth ground wire
Analyzer warm-up time:
Use this procedure to adjust the vertical position and focus of the analyzer color display.
Caution
These are the only display adjustments. Any other adjustments to the color
display will void the warranty.
Vertical Adjustment Procedure
1. Remove the left-rear bumpers and left side panel.
Figure VERT here.
Figure 3-15. Vertical Position and Focus Adjustment Locations
2. Insert the at head screw driver into the vertical position hole.
3. Adjust the control until the softkey labels are aligned with the softkeys.
Focus Adjustment Procedure
4. Use the screw driver from step 3 to adjust the focus until the display has the most
readability.
3-40 Adjustments and Correction Constants