System Info; Counters; Battery - Sunrise Medical PowerTalk2 User Manual

Table of Contents

Advertisement

9�4 System Info
Parameter

System Info

Firmware build date
Manufacturer firmware
Manufacturer serial number
Reset to factory defaults

10 Counters

10�1 Battery
Parameter
System Info
#Charge cycles
#Charge events
#BMS comm. errors
28
L
Meaning
The time and date that the system software for the main controller was
4
created. This parameter is set by the software engineer at Indes.
4
The name of the firmware developer
4
The serial number of the motor controller.
By typing "RESET_DRIVING" all the custom settings become a copy of
3
the normal settings and the default select profile is normal
By typing "FACTORY_RESET" same settings as "RESET_DRIVING"
plus below values will be set:
These counters will be set to 0: max speed, Assist and Drive rim
distances, foil errors, hall errors
And "auto off time" = 15 min, "beeps" = on, "Facility" = empty box,
"Location" = empty box, "LOG" =clear
A "FACTORY_RESET" can be used when the WheelDrive transfers to a
new end-user.
L
Meaning
Number of times the battery has been fully charged. A charge cycle is
counted when the battery is charged with a total of the design capacity
4
of the battery pack (see settings). A single charge cycle can comprise
multiple charge events.
Example:
A battery pack with a design capacity of 9 Ah is charged on day 1 with 4
Ah. On day 2 it is charged with 2 Ah and on day 3 with 3 Ah. The total of
all 3 days equals the design capacity and is counted as 1 charge cycle.
Number of charge events communicated by the battery management
4
electronics. This number is not of relevance in judgement of system
health.
This number is not relevant in judgement of system health. The value
3
can be reset by clicking the arrow symbol.
WheelDrive Parameter List - E1801_20181018