Error Indication; Error Indication R-Net; Lcd-Display - Meyra 2.322 Maintenance And Service Manual

Electric wheelchair with operating module r-net
Hide thumbs Also See for 2.322:
Table of Contents

Advertisement

ERROR INDICATION

Error indication R-Net

LCD-Display

Errors will be shown in the LCD-display (15) of the operating
module in the following error code.
(a) Shows the cause of the error.
(b) Shows the error code.
Error-Code
Cause of error
1E01
The battery charger is still connected.
2C00
Battery discharged.
1505
A magnetic brake is disengaged and the
1506
push mode activated.
3800
A motor or its cable connection is defect.
3C00
7205
Lamp short circuit, left side
7209
Lamp short circuit, right side
7206
Short circuit in cable, left side
720A
Short circuit in cable, right side
7207
No connection signal, left side
7208
No connection signal, right side
(c) Shows the location of the error.
Display
Location of error
JSM
Operating module
PM
Power module
ISM
Lighting / Adjustment module
STLM
Lighting / Steering module
DUAL
Control unit for accompanying person
(d) Shows the type of error.
Display
Type of error
Overload of the drives
System performance reduced. Self pro-
tection against overheating
c
a
c
a
15
b
b
d
43