ERROR DIAGNOSTICS

Operating module < LED >
Error illustration through the battery
gauge
In case of an error the electric travelling
wheelchair is shut down for safety reasons
and the LEDs of the battery gauge (1) blink.
The number of blink impulses indicate the
possible fault source.
☞ Note:
If the malfunction cannot be repaired
and the electric travelling wheelchair
no longer operated, contact an author-
ised specialist workshop.
Fault
1 blink impulse
2 blink impulses
3 blink impulses
4 blink impulses
5 blink impulses
Cause
The battery is discharged.
The
electrical
tion to the left motor is
open-circuit.
The electrical connection
to the left motor is defec-
tive (short-circuit).
The
electrical
tion to the right motor is
open-circuit.
The electrical connection
to the right motor is defec-
tive (short-circuit).
1
Remedy
Immediately charge the bat-
teries and if necessary check
the battery contacts.
connec-
Check the motor connection
cable, motor cable plug and
motor.
Check the motor connection
cable, motor cable plug and
motor.
connec-
Check the motor connection
cable, motor cable plug and
motor.
Check the motor connection
cable, motor cable plug and
motor.