Front Panel - UNI-T UPO3000E Series User Manual

Ultra phosphor oscilloscope
Hide thumbs Also See for UPO3000E Series:
Table of Contents

Advertisement

2.Observe the waveform displayed.
Excessive compensation Correct compensation Insufficient compensation
Picture 1-2 Probe compensation calibration
3.If the displayed waveform does not look like the above "correct compensation"
waveform, use a non-metallic screwdriver to adjust the probe' s variable
capacitance until the display matches the "correct compensation" waveform.
Warning: To avoid electric shock when measuring high voltage using
the probe, please ensure that the probe insulation is in good condition
and avoid physical contact with any metallic part of the probe.