Chapter 12 Auxiliary Function Settings

Press the UTILITY key to enter the auxiliary function settings menu.

12.1 System Function Settings

Self-correcting:
Self-correcting allows the oscilloscope to work optimally to obtain the most
accurate measurements. You can perform this function at any time, especially
when the ambient temperature range reaches or exceeds 5℃. Before
performing the self-correcting operation, make sure that the oscilloscope is
operating for more than 20 minutes.
Press UTILITY
self-correcting and a warning dialog box pops up: "Please
confirm that no signal is connected to any channel". After confirming, press
OK to start self-correcting, it takes about 5mins.
System Information:
Press UTILITY
system information to view the oscilloscope's model number,
software and hardware version number.
Language:
Press UTILITY
language to select simplified Chinese or English.
Menu Display:
The time of the menu display. You can press to show or hide the menu.
Press UTILITY
menu display, you can select 1s, 2s, 5s, 10s, 20s, or
manual by adjusting the Multipurpose knob.
Square Wave Output:
Press UTILITY
PgDn
square wave output to set the square wave
output frequency: 10Hz, 100Hz, 1kHz, 10kHz.
Frequency meter:
Press UTILITY
PgDn
frequency meter, select on and
will be displayed at the top of the screen.
The frequency meter is the counter of trigger event frequency in the trigger
channel, it is valid when trigger type is edge or pulse width.
AUX Output Selection:
Press UTILITY
PgDn
output selection to select trigger or Pass/Fail.
Trigger: AUX output terminal outputs the trigger synchronization signal.
Pass/Fail: AUX outputs the pass/fail signal.
Backlight Brightness:
Press UTILITY
PgDn
backlight brightness, and adjust the brightness
of the screen by the Multipurpose knob.
Clear Data:
Press UTILITY
PgDn
backlight brightness
stored in the device.
Network Settings:
Connect the device with a valid network cable, then press UTILITY
PgDn
IP to enter the IP setting interface.
IP Type:
Press the type key to switch the IP access mode: manual or automatic.
IP Address:
The format of the IP address is nnn.nnn.nnn.nnn, the range of the first nnn
is from 1 to 223, the range of the other three nnn is from 0 to 255. It is
recommended that you consult your network administrator for an available
IP address.
Press IP address to enter the IP address setting menu. When the access
mode is manual, set the IP address by adjusting the Multipurpose knob.
When the access mode is automatic, you can only view the IP address.
F1 to clear the data