Assembly Replacement
Front Frame Subassemblies

Front Panel External Trigger Cable

Removal/Replacement
The front panel External Trigger connector/cable must be replaced as
NOTE
an assembly.
Be careful to not scratch the dress panel when removing or replacing
CAUTION
this part.
1. Remove the front frame assembly. Refer to the
2. Refer to
3. Disconnect the cable from the A12 analog IF assembly.
4. For replacement, position the connector/cable through the front
5. Using the 9/16 socket, replace the nut to secure the connector to the
6. Clip the cable into the cable clamps on the shield.
7. Re-route the cable to avoid interference with the fans or the airflow,
8. Replace the front frame. Refer to the
Figure 4-40
Front Panel External Trigger Input Connector Removal
230
removal procedure on
Figure
4-40. Using the 9/16 socket, remove the nut (1) that
secures the connector to the front frame.
frame.
front frame. Torque to 21 inch pounds.
and reconnect to the A12 analog IF assembly.
procedure.
page
175.
"Front Frame"
"Front Frame"
replacement
Chapter 4