Replacement

1. Attach the new connector to the bracket. Torque to 120 inch pounds.
2. Attach the attenuator to the input connector. Torque to 10 inch
pounds.
3. Replace the attenuator/connector into the instrument.
4. Using the T-10 driver, replace the 2 screws that secure the input
connector bracket to the deck. Torque to 9 inch pounds.
5. Using a 5/16 inch wrench, reconnect the semi-rigid cable to the
attenuator/connector. To avoid twisting the cable, hold onto it during
tightening. Torque to 10 inch pounds.
6. Replace the front frame. Refer to the
procedure.
7. Replace the instrument outer case. Refer to the
Case"
removal procedure.
Chapter 4
Assembly Replacement
RF Input Connector
"Front Frame"
replacement
"Instrument Outer
215