Treatment screen and clinical setup menus

Parameter
AutoSet
Treatment screen
AutoSet
ü
Min
Max
APAP
Min
Max
CPAP
ü
Ramp
ü
Settings
Mode
ü
Set pressure
(CPAP)
Max pressure
ü
Min pressure
ü
Max ramp
ü
Start pressure
ü
EPR
ü
EPR level
ü
Mask
ü
Tube type
ü
20
Modes
Default
APAP
CPAP
4 cm H
20 cm H
ü
4 cm H
20 cm H
ü
ü
ü
ü
9 cm H
ü
20 cm H
ü
4 cm H
ü
45
ü
ü
minutes
4 cm H
ü
ü
Off
ü
ü
Off
ü
ü
Pillows
ü
ü
SlimLine
ü
ü
Range
O
2
O
2
O
2
O
2
O
4–20 cm H
O,
2
2
0.2 cm H
O
2
increments
O
Min–20 cm H
O,
2
2
0.2 cm H
O
2
increments
O
4–Max cm H
O,
2
2
0.2 cm H
O
2
increments
Off–45 minutes,
5-minute
increments
O
4–Min pressure
2
(AutoSet or APAP
mode), 4–Set
pressure (CPAP
mode),
0.2 cm H
O
2
increments
Off / Ramp Only /
Full Time
Off / 1 / 2 / 3 /
Patient
Pillows /
Full Face /
Nasal / Nasal Ultra
SlimLine /
Standard
Description
Shows the treatment
pressure.
Shows the treatment
pressure.
Shows the set pressure.
Orange icon shows that
device is ramping up.
Sets the therapy mode
depending on what modes are
available on the device.
Sets the fixed treatment
pressure.
Sets the upper limit of
treatment pressure.
Sets the lower limit of
treatment pressure.
Limits the ramp times the
patient may select.
Sets the pressure at the start
of ramp, up to fixed treatment
pressure (CPAP) or minimum
treatment pressure (AutoSet
or APAP mode).
Sets when EPR is to occur.
Sets the EPR value.
Selects the type of mask used
by the patient.
Selects the type of air tubing
used by the patient.