Control Mode - Hanna Instruments pH500 Series Instruction Manual

Panel-mounted, microprocessor-based ph and orp controllers
Hide thumbs Also See for pH500 Series:
Table of Contents

Advertisement

CONTROL MODE

RELAY MODES
The control mode is the normal operational mode for these
meters. During control mode pH500 and mV600 fulfill the
following main tasks:
• convert information from pH/ORP and temperature inputs
to digital values;
• control relays and generate the analog outputs as deter-
mined by the setup configuration, display alarm condition;
• RS232 management.
In addition, pH500 and mV600 can log working data
through RS232 connection. This data includes:
• pH, mV and
C measured values;
o
• last calibration data;
• setup configuration (also from PC).
The status of the meter is shown by the LED's on the right
STATUS
Control
Alarm
OFF
---
ON
OFF
ON
ON
Meter exits control mode by pressing SETUP or CAL and
confirming the password. Note
that this command generates
a temporary exit. To deactivate
the control mode definitively, set
CONTROL ENABLE to "0" (item
# 02).
Once enabled, the relays 1 and 2 can be used in four
different modes:
1) ON/OFF, high setpoint (acid dosage);
2) ON/OFF, low setpoint (base dosage);
3) proportional, low setpoint (base dosage, if available);
4) proportional, high setpoint (acid dosage, if available).
Alarm LED (green) Relay LED (yellow) Red LED
ON
ON
OFF
LEDs
OFF
ON
ON or OFF
OFF
ON or OFF
Blinking
17