Hanna Instruments BL100 Series Instruction Manual

Hanna Instruments BL100 Series Instruction Manual

Ph & orp pump controllers
Hide thumbs Also See for BL100 Series:

Advertisement

BL100 • BL101
pH & ORP Pump Controllers

Advertisement

Table of Contents
loading

Summary of Contents for Hanna Instruments BL100 Series

  • Page 1 BL100 • BL101 pH & ORP Pump Controllers...
  • Page 2 If you need additional technical information, do not hesitate to e‑mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com. All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS 1. PRELIMINARY EXAMINATION ....................4 2. SAFETY MEASURES ......................6 3. ABBREVIATIONS ........................ 6 4. SPECIFICATIONS ......................... 7 4.1. BL100 pH CONTROLLER ..................... 7 4.2. BL101 ORP CONTROLLER ................... 8 4.3. pH & ORP PROBE SPECIFICATIONS ................9 5.
  • Page 4: Preliminary Examination

    1. PRELIMINARY EXAMINATION Remove the instrument and accessories from the packaging and examine it carefully. For further assistance, please contact your local Hanna Instruments Office or email us at tech@hannainst.com. Each instrument is delivered in a cardboard box. pH model...
  • Page 5 ORP model Supplied with BL101-00, without mounting kit • HI20083 ORP/temperature probe • ORP test solution (3 pcs.) • Power connection cable • Instruction manual • Instrument & electrode quality certificates BL101-10, with in-line mounting kit • HI20083 ORP/temperature probe •...
  • Page 6: Safety Measures

    2. SAFETY MEASURES • Do not use chlorine tablets, granular chlorine or other non-liquid chlorine applications. • Do not use the pump controller on a pool utilizing electrolytic chlorine generation (salt electrolysis). • Do not add stabilizer (e.g. cyanuric acid) to the swimming pool or spa while using pump controller.
  • Page 7: Specifications

    4. SPECIFICATIONS 4.1. BL100 pH CONTROLLER 0.00 to 14.00 pH Range -5.0 to 105.0 °C (23.0 to 221.0 °F)* 0.01 pH Resolution 0.1 °C (0.1 °F) Accuracy ±0.10 pH @ 25 °C / 77 °F ±0.5 °C (±0.9 °F) • User calibration: automatic, one or two-point with buffer solution (4.01, 7.01, 10.01 pH) Calibration •...
  • Page 8: Bl101 Orp Controller

    Environment 0 to 50 °C (32 to 122 °F), max. 95% RH non-condensing Dimensions 90 x 142 x 80 mm (3.5 x 5.6 x 1.8") Weight 910 g (32 oz) Casing Wall mounted, built-in pump, IP65 rated * The range may be limited by the probe's limits. 4.2.
  • Page 9: Ph & Orp Probe Specifications

    • External switches can be attached to hold the control due to external events Digital Input (low reagent level, recirculation pump doesn't work) • Galvanic isolation Power Supply 100 - 240 Vac, 50/60 Hz Power 15 VA Consumption Environment 0 to 50 °C (32 to 122 °F), max. 95% RH non-condensing Dimensions 90 x 142 x 80 mm (3.5 x 5.6 x 1.8") Weight...
  • Page 10: Description

    5. DESCRIPTION 5.1. GENERAL DESCRIPTION & INTENDED USE BL100 & BL101 pump controllers are part of Hanna Instruments pool-line family and feature a single peristaltic dosing pump and a process electrode. BL100 accepts HI10053 pH probes. The BL101 accepts HI20083 ORP probes.
  • Page 11 BL100 measures the pH of a pool and offers automatic pH level control by adjusting the chemical dosing. BL101 measures the chlorine level of a pool and offers automatic ORP level control by adjusting the chemical dosing. BL100 BL101 can work together, stabilizing first the pH value and later the ORP value in the pool.
  • Page 12: Functional & Display Description

    5.2. FUNCTIONAL & DISPLAY DESCRIPTION BL100 Front Panel 1. Keypad area 2. CAL key – Press calibration key to enter calibration mode. 3. MENU key – Press menu key to enter setup mode and move through the menu. A long press of this key is used to exit the menu and return to measurement.
  • Page 13 BL101 Front Panel MENU BL101 BL101 ORP Pump ro e oller 1. Keypad area 2. CAL key – Press calibration key to enter calibration mode. 3. MENU key– Press menu key to enter setup mode and move through the menu. A long press of this key is used to exit the menu and return to measurement.
  • Page 14 Rear Panel & Internal View Use a Phillips head screwdriver and remove the four screws, pull back the cover and remove it. To replace the rear panel, insert the four legs back in place and tighten the four screws that secure the panel to the enclosure.
  • Page 15 Alarm Relay, Power & Level Input Normally Open ALARM RELAY OUTPUT Common SPDT 2.5 A / 230 Vac Normally Closed Pump controller not powered Alarm condition Working condition with no alarm Line POWER INPUT Protective Earth Neutral Low level sensor in tank LEVEL SENSOR...
  • Page 16 Position Description Tubing fittings Sensor input Cable gland for level input Drainage opening Cable gland for alarm relay Cable gland for power cable • Always disconnect the instrument from power when making electrical connections. • Do not access the inner panel marked Caution. •...
  • Page 17: Installation

    6. INSTALLATION General Guidelines • Select controller location so that it is shielded from direct sunlight, dripping water and excess vibrations. • Keep flow rate as constant as possible for optimum sensor operation. • Install cable gland fittings and plugs as needed, to properly seal the pump controller. •...
  • Page 18 Installation Schemes We are proposing a few typical installation schemes: in-line and flow cell. In-Line Installation, Overview & Parts Table A detailed representation of an in-line installation with relevant components is found below. Position Description Electrode Flexible PVC tubing Rigid PE tubing Aspiration filter Injector, 1/2"...
  • Page 19 Flow Cell Installation, Overview & Parts Table A detailed representation of a flow cell installation with relevant components is found below. The maximum pressure of the flow cell system is 3 atm (44 psi). Position Description Electrode Flow cell and adapter Mounting panel Flexible PVC tubing Rigid PE tubing...
  • Page 20 Mounting Recommendations for Saddle • Select required drill size. See table for dimension details. Drill Pipe • Place the upper part of the saddle (3) on top of the pipe (5) with the seal (4) placed over the hole. • Take the lower part of the saddle (6), together with inserted nuts (7) and align it under the upper part.
  • Page 21 Installing Aspiration Filter The aspiration filter is used in the reagent tank to filter and prevent debris from entering the tubing. • Cut required length of aspiration tubing (flexible) to reach between peristaltic pump inlet and aspiration filter. • Place the end of tubing on the filter. •...
  • Page 22 Flow Cell Installation In a flow cell configuration the water flows from the inlet valve to the flow cell and returns in the line via the outlet valve. To prepare the inlet and outlet valve assemblies, as illustrated in the drawing: •...
  • Page 23 Connecting the Probe to the Pump Controller (Flow Cell Configuration) • Remove the protective cap and verify if the O-ring (2) is in place. • Carefully insert the probe (1) into the flow cell adapter (3), paying attention not to damage the O-ring.
  • Page 24: Setup

    7. SETUP • Short press MENU key to enter in Setup and to move to next item in menu. • Long press MENU key to exit. • Press arrow keys to change the values. • Short press MENU key to automatically save modified values. Table below presents an overview of BL100 menu with ranges and factory set defaults.
  • Page 25 Table below presents an overview of BL101 menu with ranges and factory set defaults. Parameter Range / Option Default Settings (Scrolled message) Control Auto or oFF Auto Control type on/oFF Proportional on/oFF Hysteresis 10 to 100 mV – 50 mV Band –...
  • Page 26 Control Option: Auto or Off (oFF) to enable or disable the control With disabled option oFF, the control is off. Press one of the arrow keys for the controller settings to change from Auto to oFF and vice versa. To run a ten-seconds pump test, long press the arrow keys together until the pump starts to run. "CONTROL"...
  • Page 27 Set point General: a set point is a threshold value that will trigger control if the measurement value crosses it. With a Hi control mode, the measurement approaches the set point from a lower measurement value. With a Lo control mode, the measurement approaches the set point from a higher value than the set point. Option: user selectable Press the arrow keys to assign the set point value.
  • Page 28 Startup Delay (Automatic Control Only) Option: user selectable Startup delay represents the delay to start dosing at power-on. Press the arrow keys to change the time values. "STARTUP DELAY SEC" message is scrolled on the bottom of the LCD screen. Note: To enter Startup Delay screen, pump control mode must be set as Auto.
  • Page 29 Level Alarm Option: Enabled (En) or disabled (diS) Press the arrow keys to switch between the options. "LEVEL ALARM" message is scrolled on the bottom of the LCD screen. Note: Option can only be used with properly wired level sensor. High Alarm Option: Enabled (En) or disabled (diS) Press the arrow keys to switch between the options.
  • Page 30 Low Alarm Value Option: user selectable The range is influenced by the high-alarm value set (e.g. if high-alarm value is set as 8 pH, low-alarm value can be set from 0.00 pH and incremented up to 7.90 pH). "LOW ALARM VALUE" message is scrolled on the bottom of the LCD screen. Note: The user can set the low-alarm value with low-alarm option enabled only.
  • Page 31: Pump Control

    8. PUMP CONTROL Pump control can be enabled (automatic control) or disabled (off; manual control). See SETUP section for further details on how to enable or disable pump control. LCD backlight color indicates the pump control status: • green - automatic control or in View menu mode •...
  • Page 32: Manual Control

    Hi control mode on the BL100 Lo control mode on the BL100 On/O Control Relay on continuously Relay on continuously Regulator Regulator at the con gured ow rate at the con gured ow rate output output Hysteresis Hysteresis Measured Measured point point value...
  • Page 33: Events Management

    9. EVENTS MANAGEMENT 9.1. ALARMS Alarms can be independently enabled or disabled in SETUP. Any event that activates the alarm turns automatic control Off, the alarm relay is deactivated and the LCD backlight is blinking red. The table below illustrates the conditions that will activate the alarm and deactivate the control pump. Alarm Description Alarm condition...
  • Page 34: Warnings

    9.2. WARNINGS Two types of warnings can be independently enabled or disabled in SETUP. If any of the warnings is active, the LCD backlight turns light green. Screenshot for Warnings Description Solution warning solution Press the up / down arrows at same time to restart pump.
  • Page 35: Calibration

    10. CALIBRATION 10.1. pH CALIBRATION (BL100 ONLY) The pH electrode may be calibrated on the BL100 using an automatic two-point calibration. The electrode should be calibrated: • Before in-line or flow cell installation • Whenever the pH electrode is replaced •...
  • Page 36 • If the buffer is not recognized (either because the pH electrode has not been placed in solution or the reading is outside accepted range), "---- WRONG" message is displayed along with CAL tag blinking. • After pH 4.01 or 10.01 buffer is accepted, the "SAVE" message is displayed and the controller returns to measurement mode.
  • Page 37 • The "CAL" tag will is displayed in measurement mode after the calibration. • If the buffer is not recognized, "---- WRONG" message is displayed. It is recommended to change the solution and / or clean the electrode. • Press CAL key to exit calibration. Note: If high accuracy is required, a two-point calibration is recommended.
  • Page 38: Orp Calibration (Bl101 Only)

    • Press CAL key to confirm the value (the "SAVE" message appears for a few seconds). • Press MENU key to exit without saving and return to measurement mode (the "ESC" message is displayed for a few seconds). 10.2. ORP CALIBRATION (BL101 ONLY) ORP calibration is a single point process calibration.
  • Page 39: Ph & Orp Clear Calibration

    10.3. pH & ORP CLEAR CALIBRATION • Press CAL key and the controller enters calibration mode. • Long press MENU key and the "CLEAR" message is displayed. • The "CAL" tag is not displayed in measurement mode, indicating no calibration (until a new calibration is performed.) LONG LONG...
  • Page 40: Measurement

    11. MEASUREMENT • Power the controller. All LCD segments will be displayed for a few seconds. After initialization has been completed, the controller displays the measurement screen. • Plug the probe into the dedicated socket. Use the key to install properly. •...
  • Page 41: Error Messages

    12. ERROR MESSAGES The pump controller shows clear error messages when erroneous conditions appear and when measured values are outside the expected range. The information below provides an explanation of the errors, and recommended action to be taken. Error messages are displayed with red LED backlight. Probe is not connected.
  • Page 42: Maintenance

    13. MAINTENANCE 13.1. ELECTRODE CONDITIONING & MAINTENANCE Preparation • Remove the electrode protective cap. Do not be alarmed if any salt deposits are present, this is normal. Rinse the probe with water. • Shake the electrode down as you would do with a clinical thermometer to eliminate any air bubbles inside the glass bulb (pH electrode only).
  • Page 43 4. Starting from the left side of the pump, grab the tubing and rotate the pump manually to the right, until the tubing is removed. 5. Grease the new tubing and place it on the left side of the pump. Manually rotate the pump to the right until the tubing is on the pump.
  • Page 44: Accessories

    14. ACCESSORIES HI10053 pH/temp. HI20083 ORP/temp. probe with probe with 2 m cable, DIN 2 m cable, DIN Quick connect Quick connect BL100-302 BL100-410 Pump cover Flow cell for with screw BL100/BL101 × 2 BL100-450 10 m BL100-411 Flow cell kit for Flow cell panel Ø...
  • Page 45 BL120-263 BL120-275 Injector saddle for Injector saddle for Ø 63 mm pipe, Ø 75 mm pipe, 1/2" thread 1/2" thread BL120-300 BL120-301 Peristaltic pump Peristaltic pump tubing kit rotor BL120-401 Flow cell valve...
  • Page 46: Certification

    If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number (engraved on the bottom of the meter) and the nature of the problem.
  • Page 47 Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.
  • Page 48 World Headquarters Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com MANBL10X Printed in ROMANIA...