with the AVC module and either a DV-4
or DV-6 series gauge tube, and to set the
desired vacuum measuring units to either
Torr or mbar.

3.1 WIRING

3.1.1 Connect the 2 wire ac line cord
to the 2 position connector on the back
of the MODEL 4/6.
3.1.2 Connect the five wires coming
from the AVC module to 12 position
connector on the back of the MODEL
4/6. The connector positions are from
left to right and the description is on the
decal on the back of the MODEL 4/6.
AVC WIRE POS
Optional
1
Blue
2
White
3
Yellow
4
Gray
5
Brown
6
Green
7
8
Optional
9
Optional
10
Optional
11
Optional
12
NOTE: Be sure connectors are firmly
pushed in to the back of the MODEL 4/6

3.2 CONFIGURATION

If you ordered an AVC with a DV-4
tube the configuration should be
"torr and d-4C".
148-0598 Rev. A
DESCRIPTION
Shield (Optional)
Gnd
setpoint
input
P>SP
P<SP
+18 vdc (unreg.)
not used
P>SP (30 mA)
P<SP (30 mA)
output Gnd
0-1 vdc output
Page 5
If you ordered an AVC with a DV-6
tube the configuration should be
"torr and d-6C".
To check your current configuration:
3.2.1 Plug the AVC module into either
a DV-4 or DV-6 tube. The tube should
be at atmospheric pressure.
3.2.2 Remove the front bezel and the
front panel. Locate the black push
button in the lower right hand corner.
3.2.3 Turn the MODEL 4/6 ON (or
plug into 115vac). Since the tube is at
atmospheric pressure, the display should
read "_or_" for over range. (It may read
E__0 or some pressure if the MODEL
4/6 is set for the wrong tube.)
3.2.4 Press and release the push button.
The display will read "torr" or "hPA".
One hPa equals one millibar.
3.2.5 Press and release the push button
again within 10 sec. And the display will
read "d-4C" or "d-6C".
3.2.6 Press and release the push button
again within 10 sec. The display will
flash the set point pressure. If you know
what your set point pressure should be,
you can adjust to that pressure using the
S.P. potentiometer on top of the AVC
module. The display will remain in the
set point mode until you press and
release the push button again.
3.2.7 Press and release the push button
again and you will return to the pressure
mode and the display will read "_or_".
If the settings were all correct, the
MODEL 4/6 is ready to measure