10.0
TECHNICAL SPECIFICATIONS
THER APY SESSION
Time
40 min per session
Start Therapy
Press MAIN button on therapy display
Stop Therapy
Press and hold MAIN button
Manual Intensity
≤ 0.5 mA per step
Increase/Decrease
CONDITIONS THAT WILL TERMINATE OUTPUT
Time
40 min per session
OUTPUT
Waveform
Regulated Current or
Voltage
D. C. Component
Maximum Output Voltage
(+/-20%)
Maximum Output Current
(+/-20%)
Pulse Duration (+/- 20%)
Pulse Repetition Frequency
(+/-20%)
Pulse Pattern
LBL-5122 Rev C NOV 2019
Biphasic symmetric,
rectangular
Constant Current
None, zero net current
200 V
8 mA at 10 kΩ
300 μsec (Fixed)
150 Hz (Fixed)
Continuous, bursts tuned to
tremor frequency between
4-12 Hz
Questions about Cala Trio? Visit CalaTrio.com/HCPFAQs
POWER
Permanent rechargable battery,
Battery Type
not serviceable or replaceable
AC 50/60 Hz
Power Source
Rated Voltage: 100-240V
Max Current: 0.5 A
Fully charged battery lasts 5
Duration
therapy sessions
ELEC TRODES
Cala Trio Band with multiuse
Type
electrodes embedded
Number of Electrodes
3
Dimensions
22mm x 22mm
MEASUREMENT ACCUR ACY
Frequency
+/- 20% in the 4-12 Hz range
ENVIRONMENTAL
Operating Parameters:
Temperature Range
Relative Humidity Range
Atmospheric Pressure Range
Transport and Storage Parameters
(Cala Trio):
Temperature Range
Relative Humidity Range
Atmospheric Pressure Range
Shelf-life and Storage
Parameters (Electrodes):
Temperature Range
Relative Humidity Range
Atmospheric Pressure Range
Expected Service Life of Stimulator and Base Station: 3 years
5-40°C (41-104°F)
15-90%
700-1060 hPa
-20-45°C (-4-113°F)
≤ 90%, non-condensing
700-1060 hPa
20-27°C (68-81°F)
≤ 90%
700-1060 hPa
25