Step 7: Once Measure 1 is saved, Cala Trio will prompt you
for Measures 2 and 3
Repeat the same process for Measures 2 and 3.
Note:
If you accidentally save a poor measurement
(e.g. you were walking or talking during the "Tremor
Task"), reset Cala Trio by pressing and holding the UP
and MAIN buttons simultaneously for 3 seconds.
5.3 Setting Stimulation Intensity for Therapy
After calibrating Cala Trio to your tremor, you will set the
stimulation intensity preset.
Cala Trio automatically ramps to the intensity preset
during all subsequent therapies.
It is important to find the maximum level of therapy that
you can comfortably tolerate for 40 minutes.
Note:
You will always be able to increase or decrease the
intensity during therapy sessions as needed.
Reminder: Ensure you have dampened the entire
circumference of your wrist with ample amounts of water
before setting stimulation intensity.
To set the intensity:
Step 1: Press the MAIN button from "press main to set
intensity". Wait 3 seconds before moving to step 2
UP
button
Step 2: Press the UP button to slowly increase intensity.
You may not feel the sensation right away
LBL-5122 Rev C NOV 2019
Questions about Cala Trio? Visit CalaTrio.com/HCPFAQs
Step 3: As you continue to press the UP button, you should
feel a tingling sensation in your hand and fingers in
the areas highlighted in the images above
Note:
Tingling sensation varies depending on Cala Trio
positioning and amount of water on the wrist.
» You may feel a stinging sensation if you did not wet your
wrist sufficiently before fastening Cala Trio.
See the note below to add more water.
» You should not feel the tingling sensation in your pinky.
Additionally, if you do not feel the tingling sensation in
the fingers highlighted in step 3, see the note below to
reposition Cala Trio.
If needed, to add more water and reposition
Cala Trio:
» Press the MAIN button and wait for therapy to stop.
» Press the DOWN button to highlight 'reset intensity'
and the MAIN button to select.
» Add water to your wrist (see section 5.1, page 8) or
adjust the placement of the band (see section 5.1,
step 3, page 8).
» Repeat steps 1-3 when you are ready to set your
intensity again (see section 5.3, page 10).
Step 4: Once you find the highest intensity level that you
can comfortably tolerate for a 40-minute therapy
session, press the MAIN button to stop stimulation
Note:
Choose an intensity below a level that causes
discomfort or muscle contraction.
Step 5: Press the MAIN button to save the intensity preset
Your Cala Trio therapy is
personalized and ready to use!