7.4 How to Recalibrate Cala Trio

In order to ensure effective therapy, or if the device
is calibrated incorrectly Cala Trio may need to be
recalibrated after initial setup.
UP
button
DOWN
button
Step 1: From the time screen, press the UP and DOWN
buttons simultaneously for three seconds to enter
the Calibration Menu
Step 2: Instruct the patient to perform their prescribed
"Tremor Task" and press the MAIN button to start
the calibration. Have the patient continue their
"Tremor Task" until all dots are blue (20 secs).
The device will vibrate to indicate when the
"Tremor Task" is complete
Step 3: After calibration, press MAIN to save. If you do
not want to save the calibration, press DOWN
and then MAIN to exit
Cala Trio therapy is re-calibrated
and ready to use!
LBL-5122 Rev C NOV 2019
Questions about Cala Trio? Visit CalaTrio.com/HCPFAQs

7.5 How to Reset Cala Trio

If Cala Trio is frozen or otherwise performing in a way that
you do not expect, you may perform a reset.
Step 1: Press and hold the MAIN and UP buttons at
the same time for a few seconds until Cala Trio
displays "powering down"
UP
button
Step 2: Release the buttons and Cala Trio will
automatically restart displaying Cala Trio

7.6 How to Clean and Store Cala Trio

To clean Cala Trio Therapy components, use a disinfecting
wipe as often as once per week.
When not using therapy, charge Cala Trio overnight on the
base station. Alternatively, store Cala Trio in the included
microfiber bag at room temperature.
Note:
Do not place or store a Cala Trio band on top of the
base station unless it is attached to a Cala Trio stimulator.