Change-out
g
The HE filter must be changed if:
the low airflow alarm is activated; ;
the airflow as shown by the airflow indicator falls below 6 CFM; or
the HE filter becomes damaged or wet.
For filters with nuisance level organic vapor relief,
earlier if smell, taste or irritation is detected.
The TR-3600 should be replaced when it is dirty or damaged and
The TR 3600 should be replaced when it is dirty or damaged and
every time the HE filter is replaced.
The TR-362 should be replaced any time that any damage is noted.
An excessively loaded HE filter, prefilter or spark arrestor/prefilter
may also reduce battery pack run time. If run time is shorter than
expected replacing the filters may resolve the problem.