Specifications - Fluke 110 User Manual

True-rms multimeter
Table of Contents

Advertisement

Specifications

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %.
Extended specifications are available at www.Fluke.com.
Maximum voltage between any
terminal and earth ground ........................................ 600 V
W Fuse for A input (115 & 117 only)........................ 11 A, 1000 V, IR 17 kA
Display
Digital........................................................................ 6000 counts, updates 4/s
Bar Graph ................................................................. 33 segments, updates 32/s
Temperature
Operating .................................................................. -10 °C to 50 °C
Storage ..................................................................... -40 °C to 60 °C
Temperature Coefficient............................................ 0.1 x (specified accuracy)/°C (<18 °C or >28 °C)
Altitude
Operating .................................................................. 2000 meters
Storage ..................................................................... 10 000 meters
Relative Humidity....................................................... 95 % to 30 °C, 75 % to 40 °C, 45 % to 50 °C
Battery......................................................................... IEC 6LR61
Battery Life
113............................................................................ Alkaline: 300 hours typical, without backlight
110, 114, 115, 117.................................................... Alkaline: 400 hours typical, without backlight
Safety .......................................................................... IEC 61010-1: Pollution Degree 2
113............................................................................ Measurement CAT IV 600 V
110, 114.................................................................... Measurement CAT III 600 V
115, 117.................................................................... Measurement CAT III 600 V, 10 A
Ingress Protection ..................................................... IEC 60529: IP42 (non-operating)
Electromagnetic Compatibility (EMC)
International .............................................................. IEC 61326-1: Portable Electromagnetic Environment
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the
internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage
power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring
electromagnetic compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio
reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object.
Korea (KCC) ............................................................. Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of
it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) ............................................................... 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
IEC 61010-2-033
CISPR 11: Group 1, Class A
True-rms Multimeter
Specifications
15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

113114115117

Table of Contents