Charging your external battery
Hide thumbs Also See for Stellar 150:

Advertisement

Remember:
Once main power is restored, the Stellar™150 will automatically revert back to AC power as
long as it remains plugged in to mains power. It is a good idea to leave the device plugged in
even when you are using the external battery.
Remove the DC cable from the inlet at the back of the device, and disconnect the Hubbell
connector from the backup battery.
Be sure to connect the battery back up to the charger and make sure the charger is plugged
into an AC outlet.
Companion Guides Including:
• Care and Use of the Stellar™ 150 Internal Battery and Internal Battery Alarm
Messages
• How to Use Your External Battery: A Guide for Clients, Clinicians, and Caregivers
• Stellar™100 and 150 User Guide
• Care and Use of the Stellar™150 Internal Battery
Can be found on our website at
2020 September. Ontario VEP Stellar™150:
to Your External
Battery: ©2020 Ontario Ventilator Equipment Pool. All rights reserved. Rev date: 10/01/2020
Once connected to the external
battery, the Treatment screen will
display DC power in use with a green
illuminated DC LED.
The internal battery will be crossed
out in yellow to indicate it is not in
use.
The top of the screen will show
"External DC Power in Use!"
Push the Alarm silence button to
disable the audible alarm, the yellow
LED will stay illuminated to warn the
external battery is in use.