Date:
Mammography X-ray unit S/N:
* Measurements should be taken using a minimum 3 sec. exposure
T
a
g r
t e
F /
l i
e t
: r
M
Measured tube voltage:
28kVp
Total
Radiation
Output mGy
Repeat #
Rep #1
Rep #2
Rep #3
Mean
Criteria
≥7.0mGy air kerma/sec(800mR/sec)
Comments:
3.5 Radiation Output Rate
Table MP-4
Inspectors Name:
/ o
M
o
Measured
Radiation
exposure time
Output Rate
sec
mGy/s
79
3 .5 Radiation Output Rate
Pass/Fail
P Pass
F Fail