Performance Tests
Performance Test Procedures
Table 4-4 Recommended Test Equipment for 11974V
Instrument
Spectrum Analyzer
Synthesized Sweeper
Amplifier
Amplifier
Power Meter
Thermistor Mount
Power Sensor
Power Meter
Directional Coupler
Isolator
Millimeter-Wave
Source Module
Waveguide Straights
104
Critical Specifications
LO and IF ranges compatible with 11974V
SWEEP + TUNE voltage compatible with
11974V.
Frequency: 12.5 to 19 GHz
External leveling capability
Sub harmonics and multiples thereof: < − 25 dBc
Frequency Range: 4 to 6 GHz
Output level: +16 dBm leveled
15 dB gain from 13 to 20 GHz
Compatible with mm-wave source module
Non-harmonic spurious: < − 50 dBc
Compatible with thermistor mount
SWR: <2.0 from 50 to 75 GHz
SWR: <1.2 at 5 GHz
Compatible with power sensor
Coupling: 10 dB
Directivity: >36 dB
Primary line SWR: 1.08
Auxiliary arm SWR: 1.14
Insertion Loss: <2.0 dB
Isolation: >25 dB
SWR: <1.5
Output power: +2 dBm leveled
Output SWR: <2.0
Harmonically related spurious signals: < − 20 dBc
5 cm minimum
SWR: <1.02
Recommended
Model
8566B
w/serial prefix 2948
or later
8341B
Option 003
11975A
8349B
432A
Hughes
45774H-1100
8481A
436A
V752C
V365A
83557A
V896B