Switching Hart Revision With Field Communicator; Switching Hart Revision With Ams Device Manager; Switching Hart Revision With Loi - Emerson Rosemount 2051 Reference Manual

Pressure transmitter with hart revision 5 and 7 selectable protocol
Hide thumbs Also See for Rosemount 2051:
Table of Contents

Advertisement

Reference Manual
00809-0100-4107, Rev CA
5.7.2

Switching HART Revision with Field Communicator

From the HOME screen, enter the Fast Key sequence and follow steps within the Field Communicator to
complete the HART revision change.
From the HOME screen, enter the Fast Key sequence
Device Dashboard Fast Keys
5.7.3

Switching HART Revision with AMS Device Manager

1. Select on Manual Setup and select HART.
2. Select Change HART Revision then follow the on screen prompts.
Note
AMS Device Manager versions 10.5 or greater are compatible with HART Revision 7.
5.7.4

Switching HART Revision with LOI

Navigate to HART REV within the extended menu and select if either HART REV 5 or HART REV 7. Use
Figure 5-8
Figure 5-8. Change HART Revision with LOI
Operation and Maintenance
below to change HART Revision:
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXTENDED MENU
EXIT MENU
HART5
2, 2, 5, 2, 4
EXTENDED MENU
CALIBRAT
DAMPING
TRANSFER FUNCT
SCALED VARIAB
ASSIGN PV
TAG
ALARM SAT VALUES
PASSWORD
SIMULATE
HART REV
HART REV
BACK TO MENU
EXIT MENU
Operation and Maintenance
HART7
2, 2, 5, 2, 3
July 2017
73