Damping - Emerson Rosemount 2051 Reference Manual

Pressure transmitter with hart revision 5 and 7 selectable protocol
Hide thumbs Also See for Rosemount 2051:
Table of Contents

Advertisement

Reference Manual
00809-0100-4107, Rev CA
Figure 2-10. Analog Zero and Span Buttons
A. Zero and span buttons
If the transmitter security is on, adjustments to the zero and span will not be able to be made. Refer to
"Configure security and simulation" on page 50
The span is maintained when the 4 mA/1 V point is set. The span changes when the 20 mA/5 V point is
set. If the lower range point is set to a value that causes the upper range point to exceed the sensor
limit, the upper range point is automatically set to the sensor limit, and the span is adjusted
accordingly.
Regardless of the range points, the Rosemount 2051 will measure and report all readings within the
digital limits of the sensor. For example, if the 4 and 20 mA(1—5 Vdc) points are set to 0 and 10 inH
and the transmitter detects a pressure of 25 inH
percent of range reading.
2.6.4

Damping

The damping command changes the response time of the transmitter; higher values can smooth
variations in output readings caused by rapid input changes. Determine the appropriate damping setting
based on the necessary response time, signal stability, and other requirements of the loop dynamics
within your system. The damping command utilizes floating point configuration allowing the user to
input any damping value within 0.0—60.0 seconds.
Damping with a Field Communicator
From the HOME screen, enter the Fast Key sequence
Device Dashboard Fast Keys
Enter desired Damping Value and select APPLY.
Damping with AMS Device Manager
Right select on the device and select Configure.
Configuration
A
for security information.
O, it digitally outputs the 25 inH
2
Configuration
July 2017
O,
2
O reading and a 250
2
2, 2, 1, 1, 5
17

Advertisement

Table of Contents
loading

Table of Contents