Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

3 Disconnect the connectors of the pump assembly (J16,
J17), fan (J25) and damper (J24) at the main board and lift
the three assemblies out of the foam. Disconnect all
remaining connectors at the main board.
5 Push the leak sensor cable through the recess in the foam
and lift the bottom foam out of the instrument.
1200 Series QP Reference Manual
4 Remove the LPM board, see
Pressure Pump Main Board (LPM Board)"
6 Loosen and remove the power supply screws at the rear
panel.
Power supply screws
Repairing the Pump
"Exchanging the Low
on page 130.
149
4