Agilent Technologies 1200 Series Reference Manual page 132

Quaternary pump
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

4
Repairing the Pump
GPIB
Remote
Analog output
Screws and Nut
Figure 25
4 Remove the connector screws from the GPIB, Remote and the analog
5 Remove the board. Place the board on the ESD kit.
The RFI spring plate sitting on the board connectors is very sharp! Be careful, not to
WA R N I N G
cut yourself, when removing it from the old board an sliding it onto the new board.
6 In most cases the RFI spring plate (Radio Frequency Interference spring
If you have to exchange other parts also, continue with that work, first.
N O T E
132
Board recess
Removing the Screws at the Back Plane.
pressure output connector.
plate, prevents radio emissions from the instrument to ambient) remains on
the interface connectors of the board. Carefully remove the spring plate
from the old board and slide it onto the new board before installing the
board in the pump (the RFI spring plate is NOT part of an exchange board).
Board recess
1200 Series QP Reference Manual