THE SYSTEM REQUIRES SPECIAL PRECAUTIONS REGARDING EMC. Install and put into service according to the EMC
information provided below:
The system is intended for use in the electromagnetic environment specified below. The customer or user of the system should assure that it is used in such
an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC 61000-3-3
The system is intended for use in the electromagnetic environment specified below. The customer or user of the system should assure that it is used in such
an environment.
Immunity test
IEC 60601 Test Compliance Level
Electrostatic discharge
±8 kV contact
(ESD)
±2, ±4, ±8, ±15 kV air
IEC 61000-4-2
Electrical fast transient/
±2 kV for a.c. mains
burst
±1 kV for I/O ports
IEC 61000-4-4
Surge
± -0.5 kV, ± -1 kV, ± -2 kV line to ground
IEC 61000-4-5
± -0.5 kV, ± -1 kV line to line
Voltage dips
100% Vac for 0.5 cycles at 0°, 45°, 90°,135°,180°, 225°,
IEC 61000-4-11
270°, 315
100% Vac for 1.0 cycles at 0°
30% Vac for 30 cycles at 0°
100% Vac for 250 (50Hz) cycles or 300 (60Hz) cycles at 0°
Voltage interruptions
0% a.c. 250(50 Hz) or 300(60 Hz) at 0°
IEC 61000-4-11
Power frequency
30 A/m
(50/60 Hz) magnetic field
IEC 61000-4-8
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Class A
Not applicable
Not applicable
Guidance and manufacturer's declaration - electromagnetic immunity
Electromagnetic environment - guidance
The System uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The emission characteristics of this system make it suitable for use in
industrial areas and hospitals (CISPR 11 Class A). If the system is
used in a residential environment (for which CISPR 11 Class B is
normally required), this equipment might not offer adequate
protection to radio-frequency communication services. The user
might need to take mitigation measures, such as relocating or
reorienting the equipment.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with a synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the system requires
continuous operation during power mains interruptions, it is
recommended the system be powered from an
uninterruptible power supply or battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
19