Warnings - Bayer HealthCare MEDRAD Stellant Operation And Installation Manual

Imaging system interface (isi) 700 module ct injection system
Hide thumbs Also See for MEDRAD Stellant:
Table of Contents

Advertisement

Warnings

Procedure request can be initiated from EITHER the scanner or the injector. When either or
both system symbol (Injector and Scanner) is gray, that system is not ready and injection will
not proceed. When both the symbols are highlighted in yellow and flashing, both the systems is
ready and the injection can proceed.
Test inject is programmed, test inject can be started only from the injector. When injector
symbol is gray, injector is not ready and test inject will not proceed. When injector is highlighted
in yellow, injector is ready and the test inject can proceed.
The following symbols appear throughout this manual.
WARNING: Indicates that the information is a warning. Warnings advise you of circumstances
that could result in injury or death to the patient or operator. Read and understand the warnings
before operating the injection system.
CAUTION: Indicates that the information is a caution. Cautions advise you of circumstances
that could result in damage to the device. Read and understand the cautions before operating
the injection system.
NOTE:
Indicates that the information that follows is additional important
information or a tip that will help you recover from an error or point you
to related information within the manual.
Improper installation of the ISI module may cause it to fail. Installation must be
completed by properly trained personnel. Contact your scanner manufacturer for
configuration and installation assistance.
Patient injury could result from using improper accessories. Use only accessories and
options provided by Bayer designed for this system.
Explosion hazard: Patient injury could result from using the injection system in the presence
of flammables (such as anesthetics). Do not use the system when flammables are present.
Fire hazard: Patient injury could result from using incorrect fuses. To avoid an electrical fire,
assure the correct type of fuse is used for replacement. The fuse must be replaced with Type T,
250 V, 2 A fuse by qualified personnel only.
Shock hazard: Patient injury could result from worn cabling or unit disassembly. To avoid
exposure to potentially hazardous voltages, do not disassemble the ISI module in any way.
Worn cabling also creates hazards. If any worn or damaged cables are detected, do not use
the ISI module. Contact Bayer or your local dealer for service or replacement.
Only use the power cord supplied with the system. Do not plug the power cord into an
extension cord or multi-outlet power strip.
Patient or operator injury may result if damaged components are used. Do not use
damaged components. Visually inspect all components before use.
Electrical Shock hazard: Equipment must only be connected to a supply mains with
protective earth.
The system should not be serviced or maintained while in use with a patient.
5