Display and Control Unit (DCU):
"ISI module failure"
Certegra Workstation:
"ISI module failure"
®
MEDRAD
Stellant ISI 700 Module Operation and Installation Manual
The ISI module did not pass a self-test.
NOTE:
If the problem cannot be rectified, go to the injector setup screen
to configure ISI "Off." Manually start the injector and scanner
separately to complete the procedure. Call for service.
NOTE:
Follow the error instructions on the display and/or contact Bayer
for further support.
Schematic Diagram
Note:
Maximum rating for scanner relay contacts: 0.75 A @ 30 VDC
Minimum rating for scanner relay contacts: 10 µa @ 10mV.
Unplug the power cable to the ISI module,
then re-plug.
12