Operational Checkout - Bayer HealthCare MEDRAD Stellant Operation And Installation Manual

Imaging system interface (isi) 700 module ct injection system
Hide thumbs Also See for MEDRAD Stellant:
Table of Contents

Advertisement

Operational Checkout

®
MEDRAD
Stellant ISI 700 Module Operation and Installation Manual
If the base has a hub installed, remove it and replace it with a switch, catalog number
SWITCH301 (pictured below).
®
6.
MEDRAD
Stellant CT Injection System with Certegra
®
MEDRAD
Stellant FLEX CT Injection System with Certegra
FLEX) ONLY - The label sheet 60728249 contains two labels for each translation. Apply
one to the injector head as shown on the label sheet 60728249. Do not use the second
label. Certegra Workstation should not have a label applied to it.
®
MEDRAD
Stellant CT Injection System (Stellant Classic), skip this step. These do not
apply to the Stellant Classic.
A basic functional checkout of the ISI module should be included as part of regular
maintenance. Verifying proper operation of the ISI module will help in detection of any
problems that may not be noticed in day to day operation. The following procedure represents
a suggested series of activities which encompass typical operation of the system. Read the
following procedure carefully before beginning the checkout. If problems are detected, call your
Services Representative.
NOTE:
Any problems detected during this or any other procedure should be
corrected before using the injection system in patient procedures.
1.
Ensure that all system safety and warning labels are in place and legible.
2.
Apply power to the system.
3.
Refer to the Configuration and Details of Operation sections below.
4.
Verify that all displays and indicators are functioning properly.
®
Workstation (Stellant CWS) and
®
Workstation (Stellant
8