Scan Delay; Isi Module Cables; Power Cable Specifications; Isi Module System Messages And Descriptions - Bayer HealthCare MEDRAD Stellant Operation And Installation Manual

Imaging system interface (isi) 700 module ct injection system
Hide thumbs Also See for MEDRAD Stellant:
Table of Contents

Advertisement

Scan Delay

ISI Module Cables

Power Cable
Specifications
ISI Module System
Messages and
Descriptions
Text Box Message
Display and Control Unit (DCU):
"Scanner not ready"
Certegra Workstation:
"Imaging system not ready"
Display and Control Unit (DCU):
"Injector Disarmed by Scanner"
Certegra Workstation:
"Injector Disarmed by Scanner"
Display and Control Unit (DCU):
"ISI module not ready"
Certegra Workstation:
Not Applicable -- No message
If the injection system software version is 17A or higher, the scan delay must be programmed
from the scanner.
Maintenance: A periodic inspection of the ISI Interface Cables is a recommended to ensure
that there are no obvious defects, such as cuts or breaks in the cable, which could disrupt
signal integrity. Ensure that the connectors are properly seated.
NOTE:
Scanner specific cables are typically available from the scanner
manufacturer. Contact the scanner manufacturer for details.
The specifications required by the MEDRAD
Module relative to the power cable (plug, receptacle, and cord) are:
Operating Temperature: 60°C minimum
Receptacle Type: IEC-60320 C13
Normal Cord Voltage: 300 VAC minimum
Wire Gauge: 1.00 mm
Cord Type: IEC 60245-1, Annex A, Designation 53, or IEC 60227-1,
Annex A, Designation 53 Certified
Cord Length: 3 m maximum
The power cable must meet applicable plug, cord, and receptacle specifications including type,
voltage, current, and safety approval markings for the country in which the power cable is being
used.
The text box messages appear on the injector screen.
What it means
When the ISI module is configured "On", or
when the user attempts to arm the injector,
the system checks for the signal, "Scanner
Ready." If the scanner is not ready, this
message results.
The injector was "Armed" or injecting, and
a disarm signal was received from the
scanner.
OR
The ISI module is not communicating with
the Injector and an attempt is made to
Arm.
This message appears when two events
occur simultaneously:
(1) The user is trying to configure the ISI
module, and
(2) The injector is not communicating with
the ISI module.
®
Stellant Imaging System Interface (ISI) 700
2
minimum
Actions to be taken
Ensure the scanner is ready. Check that all
cabling between the injector, ISI module,
and scanner are connected. Ensure power
indicator on ISI module is lit.
In some cases, this can be normal
operation. However, if this message was
received in error, then check to ensure that
all cabling between the injector, ISI module,
and scanner are connected. Ensure that
the scanner is properly configured.
Check to ensure that all cabling between
the injector, ISI module, and scanner are
connected. Ensure power indicator on ISI
module is lit.