Service-01-4 Metal Power Switch Repair; Service-01-5 Rf Key Setting - Invacare Scorpius Series Service Manual

Table of Contents

Advertisement

Scorpius
2.1.3

SERVICE-01-4 Metal Power Switch Repair

Please refer to Service Procedures SERVICE-01 Steering Shroud Repair
steps as follows:
1. Find the metal power switch connector (figure on the right).
2. Reconnect the metal power switch connector.
3. If it is not effective, go to Metal Power Switch Replacement below.
Metal Power Switch Replacement:
Please refer to Service Procedures SERVICE-01 Steering Shroud Repair steps as follows:
1. Use needle-nosed pliers to unfasten the nut.
2. Remove the metal power switch nut.
3. Take out the switch and replace it with a new one.
❖ After the replacement of parts or re-installation, please follow the exact reverse sequence for reassembly.
❖ If switch function has not been restored, refer to SERVICE-01-1 Top Control Panel Repair.
• The metal power switch can be fastened in place by hand. An overly tight fastening can
cause the touch panel to be dented inwards.
• Do not put any force on the metal power switch circuit to prevent any damage that might
affect its function.
2.1.4

SERVICE-01-5 RF Key Setting

Refer to Service Procedures SERVICE-01 Steering Shroud Repair steps as follows:
1. Find the top control board setting button.
2. Press the metal switch once, and after turning on the power, the red metal switch light should be on.
3. Press the setting button once, and the metal switch red light should flash.
4. After the flashing stops, place the RF key near the metal switch. The scooter can be unlocked, and setting
has been completed.
❖ Replace the steering head top shroud.
❖ If this has not been effective, refer to SERVICE-01-1 Top Control Panel Repair.
❖ The top control board of the scooter can memorize up to four sets of RF keys.