Installation; Bath Environment - Fluke 7008 User Manual

Calibration bath
Table of Contents

Advertisement

READ SECTION 6 ENTITLED BATH USE before placing the bath
in service.
This bath is not designed to be portable. Therefore, moving the bath once it has
been installed should be kept to a minimum.
Never move a bath that is full of fluid. This action could be
extremely dangerous and could result in personal injury to the
person moving the bath.
If the bath is going to be placed in an area where it will need to be moved
frequently, a special cart can be designed to accommodate the bath making the
bath much more portable. Hart sells carts designed for these baths. However,
even with a cart the bath should not be moved full of fluid. The fluid can splash
causing injury or if the bath and cart tip, the fluid could cause damage to the
surrounding area and personal injury to personnel.
If the bath must be moved, be sure to drain the fluid to prevent any injury. The
side of the bath with the compressor is heavier than the tank side. To safely
move the bath, two people are required.

5.1 Bath Environment

The Model 7008 Bath is a precision instrument which should be located in an
appropriate environment. The location should be free of drafts, extreme
temperatures and temperature changes, dirt, etc. The surface where the bath is
placed must be level.
If used at higher temperatures where fluid vaporization is significant, use of a
fume hood is recommended.
Allow sufficient air circulation by leaving at least six inches of space between the
bath and nearby objects. Provide overhead clearance to allow for safe and easy
insertion and removal of probes for calibration.
CAUTION:
WARNING:
Chapter 5

Installation

5-1