Emerson Rosemount Analytical 1054B C Instruction Manual page 36

Conductivity microprocessor analyzer
Table of Contents

Advertisement

MODEL 1054B C
A. Two Point Calibration. (continued)
5.
Allow the sensor to acclimate to the solution. (The
temperature reading should be stable.)
6.
Press the CAL key. 15t displays briefly (if 2nd dis-
plays, press CAL again), then the Numeric
Adjustment window displays.
7.
SCROLL (
) and SHIFT (
é
sample's conductivity value at 25°C as noted in
Step 2, then ENTER into memory.
8.
Adjust the sample's temperature to the other nor-
mal temperature extreme of the process. To raise
the sample's temperature, a hot plate with stirrer is
recommended. To lower the process temperature,
place the grab sample container in an ice bath.
9.
Allow the sensor to acclimate to the solution. (The
temperature reading should be stable.)
10. Press the CAL key. 15t displays briefly (If 2nd dis-
plays, press CAL again), then the Numeric
Adjustment window displays.
11. SCROLL (
) and SHIFT (
é
sample's conductivity value 25°C as noted in Step
2, then ENTER into memory.
The analyzer will then calculate the true cell constant
and the temperature slope then return to reading con-
ductivity. The temperature slope of the process can
now be read. Press the CAL key twice. The display will
show SLP briefly then the calculated slope for the two
calibration points. Place the sensor in the process,
then remove the analyzer from hold by pressing the
HOLD key twice again.
The slope may be calculated from the following formu-
la:
(
Conductivity T
% SLOPE/°C=
Conductivity T
max
Where: Conductivity T
maximum process temperature, Conductivity T
conductivity at the lower process temperature, and the
³T is the difference between the maximum and mini-
mum process temperature.
EXAMPLE:
(
45K
35K
% SLOPE/°C=
60-50=10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
) to display the grab
ç
) to display the grab
ç
)
max
—1
X100
min
∆T
is the conductivity at the
min
is the
)
—1
X100
=2.8%/°C
B. Single Point Calibration - Slope Known. This is
the recommended procedure for the initial calibration if
the temperature slope of the process is known.
If you do not know the exact temperature slope value,
but wish to approximate it, refer to the following guide.
However, the conductivity reading may have reduced
accuracy compared to the value if the procedure in
Section A is performed.
Acids: 1.0 to 1.6% per °C
Bases: 1.8 to 2.2% per °C
Salts: 2.2 to 3.0% per °C
Water: 2.0% per °C
1.
Press the CAL key twice. The analyzer will display
SLP briefly, then show the temperature slope in
memory.
2.
SELECT to change the value. The analyzer will
display AdJ briefly, then show the Numeric Display
window.
3.
SCROLL (
) and SHIFT (
é
temperature slope for the process to be meas-
ured, then ENTER into memory.
4.
Obtain a grab sample of the process to be measured.
5.
Determine the conductivity of the sample using a
calibrated bench instrument or portable analyzer.
The instrument must be able to reference the con-
ductivity to 25°C or the solution must be measured
at 25°C. Note the reading. Insure that the analyzer
is in hold. Press the HOLD key twice and observe
the solid flag.
6.
Press the PV key once then press the SELECT
key once. Std will display followed by the Numeric
Display with digit flashing.
7.
SCROLL (
) and SHIFT (
é
ductivity value you noted in Step 5, then ENTER it
into memory.
8.
Install the sensor in the process, then remove the
analyzer from hold by pressing the HOLD key
twice.
The analyzer will calculate the true cell constant after
the initial calibration.
SECTION 5.0
START-UP AND CALIBRATION
) to display the proper
ç
) to display the con-
ç
29

Advertisement

Table of Contents
loading

Table of Contents