Alarm Setpoint - Emerson Rosemount Analytical 1054B C Instruction Manual

Conductivity microprocessor analyzer
Table of Contents

Advertisement

MODEL 1054B C
4.7 ALARM SETPOINT. The alarm setpoints should
be adjusted after completing the configuration proce-
dure outlined in Sections 4.1 to 4.6 (Refer to Figure
4-7).
1.
Press the PV key to ensure that the analyzer is
not in Set Mode.
Press the ALARM 1 or ALARM 2 key. SP1 or SP2
2.
will show briefly, followed by the Alarm 1 or Alarm
2 Setpoint.
NOTE:
If the alarm is set to OFF or FAULT
(Alarm 2 only), the analyzer will display
oFF or F 1 t respectively (refer to Section
4.2, Alarm Configuration).
3.
Press SELECT to adjust the value. The display
will acknowledge briefly with AdJ followed by the
Numeric Display with digit flashing.
ZERO
ALARM
1
ç
SELECT
F.S.
ALARM
2
Press
Press
Once
Once
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
4.
5.
6.
AdJ
Displays
Numeric
Change to
Briefly
Display
FIGURE 4-7. Alarm Setpoint
SCROLL (
) and SHIFT (
é
desired value.
ENTER value into memory.
Repeat Steps 2 to 5 for the second setpoint.
NOTE
Selection of µS/mS and decimal posi-
tions is achieved by pressing SHIFT (
until the µS/mS flag flashes, then
SCROLL (
) until the desired combina-
é
tion of decimal position and mS (quick
flashing)/µS (slow flashing) flag are dis-
played.
NOTE
Alarm logic may be changed from nor-
mally open (N.O.) to normally closed
(N.C.) by cutting circuits (W5, W7 & W9)
on the power supply PCB and adding
jumpers (W4, W6, & W8).
é
ACCESS
SP1/2
ENTER
ç
SELECT
Press
Displays
desired
Once
value
SECTION 4.0
CONFIGURATION
) to display the
ç
)
ç
Numeric
Display
Briefly
of
Setpoint
25