4.4 TEMPERATURE. Display Mnemonic t-C. Select this item
for temperature reading and compensation choices.
A. Temperature Display. Display Mnemonic d-t. Select
this item to toggle between °F and °C temperature dis-
play. The analyzer will show all temperatures in units
selected until the selection is changed.
B. Automatic Temperature Compensation. Display
Mnemonic Atc. The analyzer will use the temperature
input from the sensor for temperature correction when on
is selected. When oFF is selected, the analyzer will use the
value entered by the user for temperature correction. This
manual temperature option is useful if the temperature
sensor is faulty or not on line. Temperature specific faults
will be disabled (refer to Section 8.0).
4.4.1 Temperature Configuration t-C. Refer to Figure 4-4.
SEt
d-t
t-C
Atc
Figure 4-4. Temperature Configuration
1.
Enter Set Mode by pressing ACCESS key twice
) until t-C appears on the display.
2.
SCROLL (
é
3.
SELECT to move to the next menu level. d-t will
display.
4.
SCROLL (
) to display desired item then SELECT it.
é
5.
If d-t is selected, display will show °C or °F.
If Atc is selected, display will show on or oFF.
6.
SCROLL (
) then ENTER desired item into memory.
é
7.
If °C, °F or on are entered, display will return to the
previous level (proceed to Step 9).
If oFF is selected, numerical display will flash indicat-
ing that a process temperature value is required
(proceed to Step 8).
8.
Use SCROLL (
) and SHIFT (
é
desired value. ENTER value into memory.
9.
Repeat Steps 4-8 as required for other item.
10. Press the ENTER key to return to Set Menu.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
4.5 CURRENT OUTPUT. Display Mnemonic is OUt. This
item is used to configure the output signal.
A. Output Dampening. Display Mnemonic dPn.
Dampens the response of the signal output. This option
is useful to minimize the effect of a noisy reading. The
number entered is the sample time (in seconds) for an
averaged output. Zero to 255 seconds may be entered.
B. mA Output Range. Display Mnemonic CUr. Selection
of this item will allow choice of 0 to 20 mA or 4 to 20 mA
output range.
C. Display Output. Display Mnemonic d-O. This item is
used to select logic of output display. Selecting this item
will allow the analyzer to display current output as mA
(doc) or as a percent of full scale output range (Pct).
4.5.1 Current Output Configuration OUt. Refer to
Figure 4-5.
*F
on
*C
oFF
1.
2.
3.
4.
5.
6.
7.
8.
) to display the
ç
9.
SEt
dPn
Out
CUr
d-O
Figure 4-5. Current Output Configuration
Enter Set Mode by pressing the ACCESS key twice.
SCROLL (
) until OUt appears on the display.
é
SELECT to move to the next menu level. dPn will
display.
SCROLL (
é
) then SELECT desired item.
If dPn is selected, numerical display will flash indicat-
ing that a value is required (proceed to Step 6).
If CUr or d-O is selected, proceed to Step 7.
SCROLL (
) then SHIFT (
é
value. ENTER into memory
SCROLL (
) then ENTER desired item.
é
Repeat Steps 4-7 as required.
Press the ENTER key to return to the Set Menu.
SECTION 4.0
CONFIGURATION
420
020
doc
Pct
) to display the desired