[:SENSe] Subsystem
[:SENSe]:DATA? . . . [<data handle>]
COMMAND
Queries the current measurement result data of the :SENSe subsystem.
Valid <data handles>:
QUERY RESPONSE
COMMENTS
None
RELATED
FRONT-PANEL
KEYS
[:SENSe]:FILTer:FM:AUTO . . . <Boolean>
COMMAND
Turns the Counter's ability to automatically compensate for frequency
modulation ON or OFF. (*RST: ON)
COMMENTS
Shift + Menu (FM > AUTO | OFF)
RELATED
FRONT-PANEL
KEYS
4-24
"[SENSe:][XNONE:]FREQuency [1] | 2"
"[SENSe:][XNONE:]POWer [2]"
Frequency values are returned in Hz as ASCII bytes in NR1 format.
Power values are returned in dB in NR2 format.
Query only.
Does not initiate any measurement action.
The data handle is optional. If it is omitted, the query returns values
for all functions enabled by the FUNCtion:ON command.
Values are returned in the same order as the functions returned by
the FUNCtion:ON? query.
If this query executes while a measurement is in progress, the prior
measurement result is returned, if the prior result was not
invalidated.
When FM:AUTO is ON, the Counter automatically detects FM signals
and modifies its measurement algorithm accordingly.
Using FM:AUTO increases the time required to compute each
measurement but increases the accuracy of FM signal measurements.