ADC ADView 2 User Manual page 46

Modular diagnostic station
Table of Contents

Advertisement

Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
for 0,5 cycle
on power supply
40 % U
input lines
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the A.C. mains voltage prior to application of the test level
In the event of an error, the device will auto-recover within 5 seconds.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
3 V/m
Radiated RF
80 MHz to
IEC 61000-4-3
2.5 GHz
93-9005-00
September 1, 2016
<5 % U
T
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
T
)
(30 % dip in U
T
for 25 cycles)
<5 % U
T
T
)
(>95 % dip in U
T
for 5s
3A/m
Compliance
level
3 Vrms
3 V/m
Mains power quality should be that of a
typical commercial or hospital
)
T
environment. If a dip or an interruption of
mains power occurs, the current of the
ADView 2 may be dropped off from
normal level, it may be necessary to use
)
uninterruptible power supply or a battery.
T
)
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the ADView 2, including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
Distance in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey, should be less than the
ADView 2 User Manual | 46