ADC ADView 2 User Manual page 41

Modular diagnostic station
Table of Contents

Advertisement

3
85 & 89
86
90, 91, 97-107
505 & 506
501-504, 507-514,
516-517
520-524 & 526
530
121-124, 133-136,
150-156
126-132
137-149 & 157-192
601-610 & 612,614
611
93-9005-00
September 1, 2016
No action possible. Values are outside the reportable
Check that the hose has no sharp bends and is not pinched.
Check that the patient is not lying on the cuff.
Check that the cuff is in the correct position.
Take another BP reading.
If cancelled reading was unintended, allow cuff to deflate & restart reading.
If cancelled reading was intended, no action needed.
Power off and repower on the ADView 2.
If issue persists, call SMI Customer Service. Service Required
Temperature:
Make sure patient has not had any warm beverages recently (oral), and then retry
measurement.
Remove and then replace the probe in the well.
Retry reading.
Power off and repower on the ADView 2.
Retry reading. If issue persists, call customer service.
Power off and repower on the ADView 2.
Retry reading.
If issue persists, replace the probe. Retry reading.
If issue persists, replace module or call Customer Service.
Use in normal room temperature environment suitable to the device.
SpO
:
2
Remove sensor from the patient and ensure proper positioning. Attempt new measurement
ensuring proper protocol as described in Section 7.
Unplug sensor and reconnect.
If issue persists, power off and repower on the ADView 2.
Retry reading. If issue persists, call Customer Service.
Power off and repower on the ADView 2.
If issue persists, replace module or call Customer Service
Electronic Medical Records (EMR):
Check EMR configuration, wireless strength, and proximity. Call your biomedical engineer or
IT support.
Ensure proper barcode and patient ID scanned.
range.
ADView 2 User Manual | 41