WARNING: The ADView 2 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use
is necessary, the ADView 2 should be observed to verify normal operation in the configuration in which it will be used.
Guidance and manufacturer's declaration – electromagnetic emissions
The ADView 2 is intended for use in the electromagnetic environment specified below. The customer or the user of
the ADView 2 should assure that it is used in such an environment.
Emissions
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic
Class A
emissions IEC
61000-3-2
Voltage
Complies
fluctuations/flicker
emissions IEC
61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The ADView 2 is intended for use in the electromagnetic environment specified below. The customer or the user of
the ADView 2 should assure that it is used in such an environment. When the product is in an environment with
static, the product may show error SpO
may exhibit DEGRADATION of performance (e.g., deviation from MANUFACTURER'S specifications) that does not
affect BASIC SAFETY or ESSENTIAL PERFORMANCE.
Immunity test
IEC 60601
test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines and
patient coupled
IEC 61000-4-4
lines
Surge
±1 kV line(s) and
neutral
IEC 61000-4-5
93-9005-00
September 1, 2016
Electromagnetic environment-- guidance
The ADView 2 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The ADView 2 is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
reading, it will recover automatically when the static is gone. Though it
2
Compliance level
±4 kV contact
±4 kV air
±2 kV for power
supply lines and patient
coupled lines
±1 kV line(s) and neutral
Electromagnetic environment
--guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 40%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
ADView 2 User Manual | 45