Invacare Medley Ergo Soft Tilt User Manual page 8

Accessory
Table of Contents

Advertisement

Invacare® Soft Tilt
designated slots on the main module, place the fixation plate
from set  underneath of the bed onto the carriage-bolts to-
gether with the spring washers and tighten the lock nuts with
a 13 mm key.
4.
Place the middle plate N back onto the main module. Place first the
normal and spring washer on the bolts M, and tighten them using a
5 mm Allen key. Ensure that the bolts are well tighten and there is
no risk they become loose after some period of time.
5.
Cut all the zip ties P to release the wings, before you start the mo-
tors. Make sure new zip ties are placed in the same location when
the main module is to be removed from the bed.
Mounted beds at Low Height
WARNING!
Risk of feet squeezing
For beds assembled in a Low Height position there
is a risk of feet squeezing between the actuators
and ground.
Attaching labels
Low version only.
1.
8
Place the two feet squeezing labels on the shield plate of the actu-
ators on both sides of the bed.
Wiring
The control box and hand control is to be hung at the footboard.
Connections at the control box
Q Power supply
R Hand control
S Right motor
T Left motor
1.
Run the motor cable underneath the bed frame toward the foot
end, on both sides of the bed and fix each of the cables with zip
ties onto the frame tube. Ensure that both motor cables are tighten
with zip ties , (marked by the circles in the picture) onto the frame
tube, first zip tie near the motor and the last 2 on the frame tube
toward the foot end. Make sure the end of the zip ties, are not
facing outwards from the bed frame.
2.
Connect the hand control to cable R and connect the power cable
Q to the mains. The hand control cable is marked
clip corresponding to
on the control box cable.
3
Setup
on the cable
3
IVCMAMED221119

Advertisement

Table of Contents
loading

This manual is also suitable for:

Soft tilt

Table of Contents