Invacare ML6 User Manual

Invacare ML6 User Manual

Cylinder with post valve for oxygen system
Hide thumbs Also See for ML6:

Advertisement

Quick Links

This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
Invacare® ML6, M9 and D
Cylinder with Post Valve
For HomeFill® Oxygen System
en HF2POST6WA, HF2POST9WA,
HF2POSTDWA
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Invacare ML6

  • Page 1 Invacare® ML6, M9 and D Cylinder with Post Valve For HomeFill® Oxygen System en HF2POST6WA, HF2POST9WA, HF2POSTDWA User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
  • Page 2 All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
  • Page 3: Table Of Contents

    5.6 Storage ........22 Contents 5.7 Optional Accessories .
  • Page 4: General

    Invacare® ML6, M9 and D Cylinder with Post Valve 1 General Read the manual 1.1 Symbols No open flame Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage.
  • Page 5: Intended Use

    General 1.2 Intended Use intended to be used by or under adult supervision only after reading and understanding the instructions and warnings of this user manual. This cylinder assembly is intended to store, regulate, and deliver supplemental oxygen to patients who may have difficulty extracting Only personnel instructed and trained in the use of the HomeFill®...
  • Page 6: Limited Warranty

    Do not return products to our factory without from Invacare or a dealer, with a copy of the seller’s invoice required prior consent. C.O.D. shipments will be refused; please prepay for coverage under this warranty.
  • Page 7 THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. SOME STATES DO NOT ALLOW THE EXCLUSION OR...
  • Page 8: Component Identification

    Invacare® ML6, M9 and D Cylinder with Post Valve 1.6 Component Identification Item Description Cylinder Assembly Cylinder Conserver/Regulator Toggle Valve Cylinder Gauge Needle Cylinder Contents Gauge Flow Dial Cylinder Fillport Cylinder Fillport Cover (not shown for clarity) 1176098-B-02...
  • Page 9: Safety

    Oxygen cylinders filled by the Invacare HomeFill oxygen system shall be used for personal use only. Not to be filled for resale or use by professional users. This cylinder contains 93 ±...
  • Page 10: General Guidelines

    – If you are unable to understand the warnings, cautions professional users. or instructions, contact a healthcare professional, – This cylinder is to be used with the Invacare HomeFill dealer or technical personnel before attempting to oxygen system. This cylinder contains 93±3% oxygen use this equipment.
  • Page 11 If you use these types of cigarettes or other sources of ignition out of the room products Invacare suggests you wash your hands prior in which this cylinder is located and away from where to using the HomeFill system. If the cylinder fillport oxygen is being delivered.
  • Page 12 Invacare® ML6, M9 and D Cylinder with Post Valve DANGER! WARNING! Risk of Fire Risk of Injury or Death Dirt or contaminants on the cylinder assembly or Contaminants in the oxygen may cause injury or death. conserver can cause fire, injury or death.
  • Page 13 To avoid injury or accessories may vary by manufacturer. Invacare accessories death: such as the cylinder bag and Invacare oxygen tubing are latex – The cannula MUST be routed and secured properly. and phthalate free. Refer to the packaging on any cylinder –...
  • Page 14 Invacare® ML6, M9 and D Cylinder with Post Valve The cylinder assembly MUST be emptied of gas prior to shipping by courier or postal service. Refer to Emptying the Cylinder Assembly. Because the cylinder contains magnetic, ferrous material, it is not suitable to be in an MRI room.
  • Page 15: Setup

    Refer to the Hydrostatic use only. Not to be filled for resale or use by Testing Date section. professional users. – Filling of cylinder assemblies is performed by Invacare WARNING! HomeFill Oxygen System only. Risk of Injury –...
  • Page 16: The Hydrostatic Testing Date

    Invacare® ML6, M9 and D Cylinder with Post Valve DANGER! WARNING! Risk of Injury or Death Risk of Injury Improper filling or abuse of this cylinder assembly, or To reduce the risk of injury: failure to follow these safety instructions may cause –...
  • Page 17: Usage

    – All cylinders MUST be inspected before attempting to fill. – If the cylinder is found to be defective or does not work properly, contact an Invacare dealer or qualified technician for service or replacement. Before filling a cylinder assembly, read and follow the information...
  • Page 18 Invacare® ML6, M9 and D Cylinder with Post Valve Prior to each use, inspect product for visible damage. Refer to Attach a standard or high flow single lumen adult oxygen nasal the External Examination section. DO NOT use if any damage is cannula with a maximum length of 2.1 m (7 feet) to the product’s...
  • Page 19 – Only use the Invacare Corporation carrying bag. 11. Breathe through the nose and feel a flow of oxygen as you inhale. 12. When finished using the cylinder assembly, perform the following steps: Turn the flow dial to setting “0”.
  • Page 20: Maintenance

    Invacare® ML6, M9 and D Cylinder with Post Valve 5 Maintenance WARNING! Risk of injury or damage – The cylinder assembly MUST be disconnected from 5.1 Maintenance the HomeFill and emptied of gas prior to cleaning or disinfecting. Refer to the Emptying the Cylinder WARNING! Assembly.
  • Page 21: Cleaning

    Maintenance Examine the cylinder gauge F until the pressure is below 500 CAUTION! psig (3447 kPa). Risk of Damage – DO NOT completely empty the cylinder. Otherwise, Cylinder pressure is below 500 psig (3447 kPa) when damage to the cylinder could occur. the cylinder gauge needle is in the red zone.
  • Page 22: Disinfecting

    Invacare® ML6, M9 and D Cylinder with Post Valve 5.4 Disinfecting 5.5 Reuse Refer to the safety information in the Maintenance section WARNING! before performing this procedure. Before transferring the product to a different user, providers MUST inspect the product for damage and Follow the cleaning/disinfectant product manufacturer’s instructions...
  • Page 23: Optional Accessories

    Maintenance 5.7 Optional Accessories WARNING! Risk of Injury or Damage Use of non-Invacare accessories may result in serious injury or damage. – Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare products.
  • Page 24: After Use

    6.2 Wear and Tear Invacare reserves the right to ask for any item back that has an alleged defect in workmanship. See Warranty that shipped with the product for specific warranty information.
  • Page 25: Troubleshooting

    Troubleshooting 7 Troubleshooting 7.1 Troubleshooting If the Conserving Cylinder fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider. Problem Possible Cause Solution No flow. Flow dial at setting “0”. Adjust flow dial to prescribed setting. Cylinder empty.
  • Page 26: Technical Data

    Invacare® ML6, M9 and D Cylinder with Post Valve Environmental Temperature Range 8 Technical Data Operating Conditions –4°F to 140°F (-20°C to 60°C) Temperature: 8.1 Technical Description The Invacare Oxygen Post Valve Cylinder Assembly is an oxygen Storage Conditions -40°F to 140°F (–40°C to 60°C)
  • Page 27 Technical Data Fill Times Normal Maximum HF2POST6WA 1 hr 50 min 1 hr 25 min* 2 hr 45 min HF2POST9WA 2 hr 20 min* HF2POSTDWA 3 hr 20 min* 3 hr 50 min *With Concentrator flow rates to patient: Concentrator Model Concentrator Flow Rate Platinum 5 and Perfecto Up to 3 L/min...
  • Page 28 Invacare Corporation Manufacturer Invacare Corp. 2101 E. Lake Mary Blvd. One Invacare Way Sanford, FL 32773 Elyria, Ohio USA Tel: 407–321–5630 44036–2125 440–329–6000 800–333–6900 Technical Services 440–329–6593 800–832–4707 www.invacare.com Canada 570 Matheson Blvd E Unit 8 Mississauga, Ontario L4Z 4G4 Canada 800–668–5324...