Invacare® Soft Tilt
Immunity test
Test / Compliance level
< 0% U
Voltage dips, short
45° steps
interruptions and
voltage variations on
0% U
power supply input
lines
70% U
IEC 61000-4-11
< 5% U
Power frequency
(50/60 Hz) magnetic
field
30 A/m
IEC 61000-4-8
Conducted RF
3V
150 kHz to 80 Mhz
IEC 61000-4-6
6V
in ISM & amateur radio
bands
10 V/m
80 Mhz to 2,7 GHz
Radiated RF
IEC 61000-4-3
385 MHz - 5785 MHz test
specifications
for immunity to RF wireless
communication equipment
refer to table 9 of IEC
60601-1-2:2014
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
10. 3.1
Test specifications for immunity to RF wireless communications equipment
IEC 60601-1-2:2014 — Table 9
Test Frequency
Band
a)
(MHz)
385
380 - 390
450
430 - 470
710
745
704 - 780
780
810
870
800 - 960
930
24
Electromagnetic environment – guidance
for 0,5 cycle at
T
Mains power quality should be that of a typical commercial or hospital environment.
If the user of this product requires continued operation during power mains interrup-
for 1 cycles
tions, it is recommended that the product is powered from an un-interruptible power
T
supply or a battery.
for 25 / 30 cycles
T
U
for 250 / 300 cycles
T
Power frequency magnetic fields should be at levels characteristic of a typical location
in a typical commercial or hospital environment.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the elec-
tromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which this prod-
uct is used exceeds the applicable RF compliance level above, this product should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating this product.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Portable and mobile RF communications equipment should be used no closer than 30
cm to any part of this product including cables.
(MHz)
Service
a)
TETRA 400
GMRS 460,
FRS 460
LTE Band 13,17
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
is the a. c. mains voltage prior to application of the test level.
T
Modulation
b)
Pulse modulation
b)
18 Hz
FM
c)
± 5 kHz
deviation
1 kHz sine
Pulse modulation
b)
217 Hz
Pulse
modulation
b)
18 Hz
Maximum
Distance (m)
power (W)
1.8
0.3
2
0.3
0.2
0.3
2
0.3
Immunity test level
(V/m)
27
28
9
28
IVCMAMED221119