Troubleshooting; Performance Troubleshooting - Invacare Modulite User Manual

Supplement to power wheelchair, seating system
Hide thumbs Also See for Modulite:
Table of Contents

Advertisement

Invacare® Modulite

8 Troubleshooting

8.1 Performance Troubleshooting

For additional troubleshooting information regarding the power wheelchair & electronics, refer to the Troubleshooting
section of the power wheelchair and remote user manuals (provided separately).
Symptom
Wheelchair power is ON, but system
does not drive
Seating system not functioning
70
Probable cause
System tilted and/or elevated beyond
the drive lockout (DLO) angle
Drive motors not engaged
LNX powered center-mounted legrest
with telescoping footboard lowered
Low batteries
Loose/faulty electrical connection
Blown fuse
Interference/obstructions, pinched wires
Solutions
Return seating system to neutral (home)
position.
Engage drive motors.
Retract footboard to top position.
Check/charge/replace batteries.
Contact your provider.
Check cable connections/check cable ties
(too tight/too loose).
Contact your provider.
Inspect/replace fuse.
Contact your provider.
Check for sources of interference or
obstructions/inspect cables for pinch
points.
Contact your provider.
1656515-B