Voltage Measurement Accuracy
To verify voltage measurement accuracy of the Live Circuit Warning function,
apply the voltages listed in Table 6 to the + and - terminals of the UUT.
Verify:
•
UUT reading is within the display limits of Table 6.
•
UUT is beeping at a 1-second interval.
•
is flashing on the display.
Voltage Source
Step
Output
1
-38 V dc
2
240 V ac, 60 Hz
Adjustment Procedure
The product should be performance tested yearly to ensure compliance with its
specifications. When required, use the following adjustment procedure to bring
the UUT within its nominal accuracy specifications.
Interface Connection
Perform adjustment with software using a computer and IR (infrared) adapter.
Connect the Infrared Cable Assembly to the UUT IR Port and COM port of the
computer. Refer to Figure 3.
1
IR Port
IR Device
Table 7. Voltage Measurement Test
UUT Display
UUT Tone
Flashing Hazard
Beeps
Flashing Hazard
Beeps
Figure 2. IR Port
Insulation Tester
Adjustment Procedure
UUT Display Limits
Minimum
Maximum
30 V
46 V
202 V
278 V