Specifications; General Specifications - Fluke 1550C Calibration Manual

Insulation tester
Hide thumbs Also See for 1550C:
Table of Contents

Advertisement

1550C/1555
Calibration Manual

Specifications

General Specifications

Display
Power
Charger Input (AC)
Dimensions (H x W x L)
Weight
Tamper Protection
Temperature (operating)
Temperature (storage)
Relative Humidity
Altitude
IP Rating
Input Overload Protection
Safety
Electromagnetic Compatibility
(EMC)
International
Korea (KCC)
USA (FCC)
Wireless Radio with Adapter
Typical Battery Charge
Capacity
At temperature extremes, the
battery needs to be charged more
frequently.
20
475 mm x 105 mm
12 V lead-acid rechargeable battery 2.6 Ahr
85 V to 250 V ac, 50/60 Hz, 20 VA
This Class II (double insulated) instrument is supplied with a Class 1 (grounded)
power cord. The protective earth terminal (ground pin) is not connected internally.
The extra pin is for added plug retention only.
170 mm x 242 mm x 330 mm (6.7 in. x 9.5 in. x 13.0 in.)
3.6 kg (7.94 lb)
Kensington lock
-20
°
C to 50
°
-20
C to 65
80 % to 31
2000 m
IEC 60529: IP40
1000 V ac
IEC 61010-1: 600 V CAT IV / 1000 V CAT III, Pollution Degree 2
IEC 61326-1: Portable
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-
coupled radio frequency energy that is necessary for the internal function of the
equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic
and those directly connected to a low-voltage power supply network that
supplies buildings used for domestic purposes. There may be potential
difficulties in ensuring electromagnetic compatibility in other environments due
to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and
may not provide adequate protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the
equipment is connected to a test object.
Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave
equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and not to
be used in homes.
47 CFR 15 subpart B. This product is considered an exempt device per clause
15.103.
Frequency Range: 2412 MHz to 2462 MHz
Output Power: <100 mW
Test Voltages
250 V
500 V
Note
1 kV
2.5 kV
5 kV
10 kV
°
C (-4
°
F to 122
°
F)
°
°
°
C (-4
F to 149
F)
°
C decreasing linearly to 50 % at 50
Number of Tests
°
C
4100
3600
3200
2500
1000
500