1550C/1555
Calibration Manual
Voltage Measurement Accuracy
To verify voltage measurement accuracy of the Live Circuit Warning function, apply the
voltages listed in Table 6 to the + and - terminals of the UUT. Verify that:
•
•
•
Step
1
2
Adjustment Procedure
The product should be performance tested yearly to ensure compliance with its
specifications. When required, use the following adjustment procedure to bring the UUT
within its nominal accuracy specifications.
Interface Connection
Perform adjustment with software using a computer and IR (infrared) adapter.
Connect the Infrared Cable Assembly to the UUT IR Port and COM port of the computer.
Refer to Figure 3.
Item
12
The UUT reading is within the display limits of Table 6.
The UUT is beeping at a 1-second interval.
is flashing on the display.
Table 6. Voltage Measurement Test
Voltage Source
UUT Display
Output
-34 Vdc
Flashing Hazard
240 Vac, 60 Hz
Flashing Hazard
1
IR Port
IR Device
UUT Tone
Beeps
Beeps
Description
Figure 3. IR Port
UUT Display Limits
Minimum
Maximum
29 V
39 V
204 V
276 V
2
ghh04.eps