Siemens SIPROCESS GA700 Operating Instructions Manual page 74

Continuous gas analysis, wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
11.2 OXYMAT 7 technical specifications
Gas inlet conditions
Correction of the internal pressure sensor
Pressure loss between sample gas inlet and
sample gas outlet
Sample gas flow
Sample gas temperature and humidity
Temperature of measuring chamber and gas path
Gas path
Standard version
74
Sample gas pressure
Devices with tubes
Devices with pipes
Without vibration compensation
With vibration compensation
Devices with tubes
Devices with pipes
Reference gas pressure
High-pressure connection
Low-pressure connection with external
reference gas pump (only for sample gas
pressure 700 to 1100 hPa (absolute))
500 to 1500 hPa (absolute)
500 to 3000 hPa (abs.); short-term max.
5000 hPa (abs.)
500 to 2500 hPa (abs.); short-term max.
5000 hPa (abs.)
500 to 1450 hPa (absolute)
500 to 2450 hPa (absolute)
Without vibration
compensation
With vibration
compensation
100 hPa above the sample gas pressure
150 hPa (with vibration compensation)
< 100 hPa at 1 l/min
18 to 60 l/h (0.3 to 1 l/min)
0 to 60 °C with rel. humidity < 90%, relative to the
respective ambient temperature of the device,
condensation in the gas path must be avoided
5 K above ambient temperature
Approx. 72 °C
Operating Instructions, 01/2013, A5E31930403-01
2000 to 3500 hPa
above the sample gas
pressure, but a
maximum of 5000 hPa
(absolute)
2500 to 4000 hPa
above the sample gas
pressure, but a
maximum of 5000 hPa
(absolute)
Wall-mounted device