10.6

Replacing the cable gland set

Replacing the cable gland set
1. Isolate the device from power.
2. Open the door. To do this, unscrew the six screws.
3. Remove the cover plate
4. Remove the entire cable gland set by loosening the screw
5. Install the complete new cable gland set
Wall-mounted device
Operating Instructions, 01/2013, A5E31930403-01
the side wall and the 11 nuts in the bottom of the housing.
(Retighten the screw
Nm.
Note
Tightening the cable glands
Fasten the cable glands
• M16x1.5 clamping range 5 to 10 mm made of plastic, counternut and cap nut
(clamping of cables or plugs) with a torque of 3.0 Nm.
• M20x1.5 clamping range 10 to 14 mm made of plastic, counternut and cap nut with a
torque of 4.5 Nm.
• M16x1.5 clamping range 6 to 10 mm made of metal, counternut and cap nut with a
torque of 6.5 Nm.
• M20x1.5 clamping range 10 to 14 mm made of metal, counternut and cap nut with a
torque of 9.0 Nm.
by loosening the three screws
: tightening torque: 2.0 Nm.)
with eleven hex nuts. Tightening torque: 0.8
Maintenance and servicing
10.6 Replacing the cable gland set
.
on the right on the inside of
67