Commissioning

9.1 General information

9.1
General information
Tighten the screw caps at the gas connection glands with a suitable open-ended wrench. In
the process, make sure that you counter properly. Otherwise, there is a danger that the gas
path will leak.
Sample gas and reference gas quality in an internal gas preparation system
Note
Gas preparation
Insufficient gas preparation can lead to contamination of the analyzer chamber. This can
result in measured value drift and temperature-related measuring errors.
● The sample gas/reference gas must be free from dust and condensation. The dust filter
● Connect the corresponding absorption filter before the device to filter out components
● Make sure that the sample gas does not condense in the gas preparation unit and in the
● The following devices can be installed upstream of the sample gas inlet for gas
Depending on the composition of the sample gas, you may need additional equipment such
as a wash bottle, additional filters and pressure reducers.
Make sure that the gas sampling devices, gas cooling devices, condensate vessels, filters
and any connected controllers, recorders or indicators are ready for operation.
52
WARNING
must permanently withhold dust particles of > 2 µm.
which are corrosive or could interfere with measurement.
analyzer itself. Appropriate measures include heated sample gas lines, a heated
analyzer, or a compressor cooler, for example.
preparation:
– A gas sampling device
– A sample gas cooler
– A filter
– A gas suction pump